Deutsch   English  

Technical catalogue
Send inquiry | back
Order number: 92-4A.3


Leakage current tester 92-4A
single-phase, non-floating

  • Leakage current measurement 0 .. 0.01;1;10mA
  • Test voltage: Grid .. 280V / 15 A internal
  • Expansion modules for a wide range of standards



Leakage current test
• Measurement range: 0-0,1/1/10 mA
• Measuring voltage internal/external: supplied mains voltage -280V / 50 -280 V
• Voltage manually adjustable: erdgebunden
• Measuring current internal/external: 15 / 15 A
• Frequency internal/external: Net frequency / 50-400 Hz
• Norm concurring
• Limit value programmable
• Measurement accuracy: At 49-61Hz: TRMS, MAD and alternating component: +/-1% rdg +10 Digit, DC component, peak value measuring and measuring to EN 60990 Illust. 3: +/2% rdg +30 Digit


Display / Operation
• menu-assisted LCD graphic display, 240x128 dots, with backlight
• programming by means of rotary encoder with pressure point
• 200 program positions are available for the storage of individual parameter blocks
• switchable to national language


Interfaces
• RS 232C-interface
• external input


Important options (optional)
• electrical outlet for other countries (optional)
• Automatical voltage setting (optional)
• Expansion module medicine technology (optional)
• Measuring circuits for: EN 60598-1 / VDE 0711, EN 60601 / VDE 0750, EN 60950 / VDE 0805, EN 60065 / VDE 0860, EN 61010 / VDE 0411, EN 60335-1 / VDE 0700 (optional)


Equipment data
• Mains voltage: 0 - 230 +10% -5% (49 .. 61 Hz)
• Safety class: I
• Slide in technique: 19"
• Depth/ Width/ Height: B = 483, T = 360, H = 177 mm
• Height : 4 HU


Order data

• Order description: Leakage current tester 92-4A
• Order number: 92-4A.3

The 92-4A serves for the norm concurring measuring and evaluation of leakage currents. Due to the integrated computer system the device can be used as an independent measuring system or as a component in an automated device tester. The on-wirings and test runs demanded in the different norms are already stored in the device and must therefore not be realized and programmed any more. The adaptation to the different norms is made by the completion with the respective measuring circuit modules. Up to three different measuring circuits can be used at the same time.



Norm conformity:

The 92-4A supports standard-compliant tests according to following norms::
EN 50106/ VDE 0700 Part 500; EN 50144-1/ VDE 0740 Part 1; EN 60065/ VDE 0860; EN 60204-1/ VDE 0113 Part 1; EN 60335-1/ VDE 0700 Part 1; EN 60598-1/ VDE 0711 Part 1; EN 60601-1/ VDE 0750 Part 1; VDE 0751-1/ VDE 0750 Part 1; VDE 0701-1/ VDE 0701 Part 1; EN 61558-1/ VDE 0570 Part 1; EN 61029-1/ VDE 0740 Part 500; EN 61010-1/ VDE 0411 Part 1; EN 60950-1/ VDE 0805 Part 1;